http://www.prometron.cz/

     
 
pH / CONDUCTIVITY / SALINITY METER CPC-505

 

• Laboratory meter in benchtop housing, powered by 12V power adapter.
• Meter is equipped with large easy-to-read backlight LCD display with brightness control.
• Measures pH and mV (redox), conductivity, salinity, temperature.
• The pH measurement circuit  is isolated from the conductivity measurement circuit.
• Unification of the operating in all measuring functions makes working trouble free.
 
Characteristics of the pH measurement:
• Calibration: 1÷5 points.
• Memory of 3 electrodes characteristics makes changing them easy.
• Automatic buffers detection, their value may be changed by the user.
• Automatic correction of the temperature influence on sample solution (NIST) value.
• Automatic checking of the condition of electrode’s membrane.
• Automatic or manual temperature compensation.
• Precise ORP (redox) measurement (accuracy 0.1 mV).
 
Characteristics of the conductivity measurement:
• Thanks to a wide conductivity measuring range they may be used for both ultra pure water and very salty samples measurements.
• 6 subranges switched automatically.
• Calibration by entering the constant K in range 0.01÷19.999 or in sample solutions.
• Works with conductivity cells equipped with platinum electrodes.
• Storing of 3 constant K of cells which cover the whole measuring range.
• Wide range of α coefficient (0÷10% / °C).
• Possibility of changing the reference temperature.
• Counting the salinity according to real characteristics for NaCl and KCl, what greatly increases the accuracy.
• Possibility of defining the TDS with use of conductivity measurement by entering the TDS coefficient in range 0.2÷1.0.
 
• Internal clock with date.
• Internal datalogger enables storing 950 measurements taken as series or single with temperature, time and date.
• USB output.
• The meters meets the GLP requirements.